TM 11-6625-2638-30
Table 2-3. Test Set Troubleshooting - Continued
STEP
PROCEDURE
NORMAL INDICATION
ABNORMAL INDICATION
CORRECTIVE ACTION
7
Press LAMP TEST switch.
All lamps shall be on.
One or more lamps off (but not all off).
Replace defective lamps.
One or more lamps still off.
All lamps off except AC and DC POWER.
Check 5 vdc per step 6 a.
8
Release LAMP TEST switch
All lamps off except AC and DC POWER. TFT lamp may be on or off.
All lamps on.
9
Connect return () lead of DVM (DC mode) to J4-h.
10
Measure the power supply voltages at the following connector pins:
4.9 to 5.5 vdc
J3-D
4.9 to 5.5 vdc
RCVR Current + jack
-4.9 to -6.1 vdc
J3-C
13.3 to 14.7 vdc
J4-d
-14.0 to 16.0 vdc
J3-L
19.0 to 25 vdc
J3-K
1 9.0 to -25 vdc
J4-E
26.0 to 30.0 vdc
J1-h
13.0 to 15.0 vdc
J5-b
(DVM to AC mode)
Move DVM Return () to J4-q
26.0 to 34.0 vac
J4-n
26.0 to 34.0 vac
J4-p
One or more voltage is higher than maximum.
All voltages missing.
Check relay K1 and interlock S9.
One or more voltages missing but not all.
11
Verify that the RCVR CURRENT switch is in the + 5V position.
Do not let the 5 vdc probe short to anything to prevent damage to power sup-
ply or test set.
12
Connect one end of a test lead with probe to RCVR CURRENT + jack (5 vdc).
2-7